参数\型号
|
BHC-1300 ⅡA/B2
|
BHC-1800 ⅡA/B2
|
洁净等级Cleanliness
|
ISO 5 (10级Class 100) / ISO 4 (10级Class 10)
|
洁净等级Cleanliness
|
HEPA/ULPA @0.3-0.12 um 99.995%-99.999%
|
过滤器级别/过滤效率Filtration Efficiency
|
1300m3h±10%
|
0.32±0.025m/s
|
洁净等级Cleanliness
|
1700m3h±10%
|
0.27±0.025m/s
|
过滤器级别/过滤效率Filtration Efficiency
|
≥0.3m/s±0.25
|
0.28±0.025m/s
|
平均送风风量Aver. Supply Air Flow (m3/h)
|
≥0.55m/s
|
平均排风风量Aver. Exhaust Air Flow (m3/h)
|
≤62dB(A)
|
平均送风风速Aver. Supply Air Velocity
|
≤5μm
|
工作窗口吸入风速Velocity inhaled in front
|
AC,单相220V/50Hz (AC, 220V,1¢,50Hz)
|
气流平衡生物防护Protectiontesting
|
人员防护Operator protection 5-8×10 CFU/ml(连续3次3 consecutive times)
|
a.撞击采样器的菌落总数≤10CFU/每次Total
|
b.狭缝式采样器菌落总数≤5CFU/每次Total
|
受试产品防护Tested product protection 5-8×10 CFU/ml(连续3次3 consecutive times)
|
菌落总数≤5CFU/每次Total
|
交叉感染防护Cross contamination 5-8×10 CFU/ml(连续3次3 consecutive times)
|
菌落总数≤2CFU/每次Total
|
最大功率Max. Consumption (KVA)
|
≤1.5kw (含备用插座 with spare socket)
|
≤1.5kw (含备用插座with spare socket)
|
重量Mass (kg)
|
380
|
450
|
工作区尺寸(宽×深×高)Work Dimension (W*D*N)
|
1300×610×640mm
|
1800×610×640mm
|
装置外形尺寸Overal Dimension (mm)
|
1500×825×2170
|
1800×820×2210
|
荧光灯规格及数量Size and Qty of fluorescent lamp
|
30W×①
|
40W×①
|
紫外灯规格及数量Size and Qty of UV lamp
|
30W×①
|
30W×①
|
照度Lumin.
|
≥750Lx
|
送、排风管口径Port Size of suppy & exhaust
|
DN250
|